Muster – Memmert PM 200/400/500 Pump Module (Generation 2003) User Manual

Page 31

background image

VO

Muster

page 31

Language
Settings: GERMAN, English, franCEAIS, ESPANOL and italIANO

Vent connection select IN1 or IN2 (option)

Settings: IN1 or IN2
e.g. IN1 for fresh air and IN2 for inert gas

Calibration correction for user-calibration of temperature.
CAL1

to CAL3

ADJUST – TEMPERATURE CALIBRATION

READJUST – TEMPERATURE CORRECTION

(see Section: „Calibration“)

Calibration correction for user-calibration of vacuum.
CAL1

to CAL3

ADJUST – VACUUM CALIBRATION

READJUST – VACUUM CORRECTION

(see Section „Calibration“)

Exit setup mode
= store all settings and exit SETUP mode.

22.1 Real-time clock

The real-time clock is set in SETUP and includes date and clock time.
The real-time clock serves for documentation according to GLP.
Date and clock time are marked in the report print.
On the graphics print the time axis is marked in real-time.
The clock runs with a buffer battery independently of the mains power supply.
The built-in lithium battery Type CR 2032 has a life of approx. 10 years.