7 saving and loading tempering profiles – Memmert CELSIUS 10 FDA User Manual
Page 42
42
CELSIUS 10 FDA-Edition
8.1.7 Saving and loading tempering profiles
In order to prevent any alteration of files, CELSIUS FDA encrypts all profile and protocol
files. Encryption employs a key which is specified in the licence file. Existing files from a
previous CELSIUS version can not be read or exported. If an encrypted file has been altered
(except through CELSIUS), access to the data is no longer possible, even for the producer
of the software.
A tempering protocol can be saved
►
To File (
or "File"
→"Save Tempering Profile"→"To File"). In the window that opens, you
can specify the file name and the storage location. In the preset directory there is a folder
each for tempering profiles and for protocols.
With "File"
→"Load Tempering Profile"→"From File" or
you can open saved tempering
profile files (*.cel) and edit them.
►
In the oven (
or "File"
→"Save Tempering Profile"→"To Device"). The tempering pro-
file is saved in the ring protocol memory of the oven. It can be opened and edited with
"File"
→"Load Tempering Profile"→"From Device".
►
To Internal MemoryCard (
or "File"
→"Save Tempering
Profile"
→"To Internal MemoryCard"). With this command,
the tempering profile on the MEMoryCard can be stored
in the controller of the oven. In this case, the
MEMoryCard must be compatible with the oven model.
You may enter a name for the tempering profile saved on
the card (max. 8 characters, no spaces) or use the name
suggested.
If you enable "Write Protection", you ensure that the tempering profile on the MEMoryCard
cannot be altered by the controller of the oven. This write protection has no effect on
CELSIUS writing data to the MEMoryCard. The default setting for write protection is ON.
Only one tempering profile can be written on a MEMoryCard XL. The MEMoryCard XL can
be overwritten at any time, however.
►
To External MemoryCard ("File"
→"Save Tempering Profile"→"To External MemoryCard (USB/
RS 232)". This allows you to save a tempering profile of up to 40 ramps on a card in a con-
nected and logged on card reader (see page 45 ).
With "File"
→"Load Tempering Profile"→"From Internal MemoryCard"/"... external" you can open
tempering profiles stored internally or externally on the MEMoryCard.