Memmert CELSIUS 10 FDA User Manual
Page 36

36
CELSIUS 10 FDA-Edition
If a coordinate is created in a prohibited area or you want to
move it to such an area, the mouse cursor turns into a
no-entry sign. The coordinate jumps automatically back to its
previous position when you let go.
Changes to the graphical view are automatically updated
when you switch to the tabular view, and vice versa.
Ramp close statements
Ramp close statements
Each segment (ramp) of a tempering profile must
end with a close statement and be linked to the next
segment. These positions are represented by the blue
dots.
To run the selection dialog, move the
mouse cursor to one of the blue ramp end points and
press the right mouse button. The following ramp
close statements are available:
►
normal:
Resets the end point definition back to the standard value. At the
end of the ramp segment the oven continues straightaway with
the next ramp segment. If this is the final ramp segment, the
tempering profile is stopped and the connected oven is reset to
its default configuration.
►
spwt()
(= Setpoint wait). The oven waits for the end of the ramp seg-
ment until the required setpoint has approximately been reached.
Only then is the next segment processed. This command cannot
be applied for the final ramp segment of a tempering profile.
Which setpoint is involved depends on the oven, and can be
seen from the expression in brackets: spwt(t) = Temperature,
spwt(h) = Humidity, spwt(p) = Pressure etc.
Temperature for incubators (Ixx, INCO 2, IPP, ICP) and HCP humid-
Temperature for incubators (Ixx, INCO 2, IPP, ICP) and HCP humid-
ity chambers:
ity chambers:
If this function is enabled, the next ramp only begins when the
setpoint temperature has been reached with a tolerance of
± 0.5° C.
Temperature for universal ovens, sterilisers and
Temperature for universal ovens, sterilisers and vacuum ovens:
If this function is enabled, the next ramp only begins when the
setpoint temperature has been reached with a tolerance of
± 2.0° C.
Vacuum for VO vacuum ovens:
If this function is enabled, the next ramp only begins when the
setpoint vacuum has been reached with a tolerance of ± 3 mbar.
For VO vacuum ovens, the SPWT function can be enabled sepa-
rately for temperature and vacuum.
Humidity for HCP humidity chambers:
Humidity for HCP humidity chambers:
If this function is enabled, the next ramp only begins when
the setpoint humidity has been reached with a tolerance of
± 2 % rh. For HCP humidity chambers, the SPWT function can be
enabled separately for temperature and humidity.
loop count
delete