4 connecting an external card reader, Logging on ovens, 1 logging on online and offline – Memmert CELSIUS 10 FDA User Manual
Page 31

31
CELSIUS 10 FDA-Edition
6.4 Connecting an external card reader
If you want to connect an external card reader (can be
acquired separately as an accessory) to your computer
so that you can store data on a MEMoryCard, the card
reader must first be logged on.
1. Connect the card reader to a USB or COM interface
of the computer.
2. In the "Settings" menu, click on the menu item
"Options".
3. Select the interface to which you have connected
the card reader (USB or COM) and close the op-
tions and settings windows with "OK". The card
reader can now be used.
If you click on "Format MEMoryCard", the card is
re-formatted. All data stored on it is then deleted.
7. Logging on ovens
7.1 Logging on online and offline
When logging an oven on, read the infor-
mation and notes concerning this in the
oven's user manual.
In order to create and edit a tempering profile, an oven must always be logged on. There are
basically two ways of doing this:
►
Logging on online:
Logging on online:
An oven can be logged on that is switched on and connected to an interface of the PC.
Only for an online log-on can the oven be controlled from the PC or the programme be
saved on a MEMoryCard (XL) located in the PC.
►
Logging on offline
Logging on offline:
As an alternative to logging on online, a virtual oven can also be logged on, which does
not have to be currently connected to the PC. This so-called offline operation is useful, for
example, when
•
the tempering profile is created initially on a notebook computer that is to be con-
nected to the oven to control it at a later time;
•
a new tempering profile is to be created while the oven is controlled from the same PC
with a different tempering profile;
•
the tempering profile is created on the PC and is to be stored afterwards on a
MEMoryCard (XL) using an external card reader.