Memmert CELSIUS 10.0 User Manual
Page 36

36
CELSIUS 10.0
When a tempering profile is started, the start dialog appears automatically with the following
setting options:
1 2
3 4 5
6 7
8
9
10
1 Log and print GLP data (automatically enabed if GLP data were preset, see page 9 )
2
Select the storage location of the log file if " Create Log File" was clicked
3 Click on this if a log file is to be generated
4 GLP description of the procedure
5 GLP value company
6 Starting time; the default is "today" and the current time; both values can be changed. The
tempering profile then starts at the set time.
7 GLP value tested by
8 If "Protocol only" is clicked, no setpoint values are sent to the oven and only the actual val-
ues are logged. The setpoint values must be set on the oven. The "Protocol only" operating
mode is possible both in the main switch position "I" and in the "clock" position of the oven.
9
Starting time
10 GLP value sample ID
6.1.7 Saving and loading tempering profiles
A tempering protocol can be saved
►
To File (
or "File"→" Save Tempering Profile"→"To File"). In the window that opens, you
can specify the file name and the storage location. In the preset directory there is a folder
each for tempering profiles and for protocols.
With "File"→"Load Tempering Profile"→"From File" or
you can open saved tempering
profile files (*.cel) and edit them.
►
In the oven (
or "File"→" Save Tempering Profile"→"To Device"). The tempering pro-
file is saved in the ring protocol memory of the oven. It can be opened and edited with
"File"→"Load Tempering Profile"→"From Device".