Memmert CELSIUS 10.0 User Manual
Page 35

35
CELSIUS 10.0
Inserting and deleting lines
If the table window is active, the commands for deleting and inserting lines are additionally
available in the view menu and toolbar:
►
Insert a line at the position marked
►
Delete selected line
If the last ramp segment is deleted, the ramp close statement is automatically transferred
to the line above.
6.1.6 Executing a tempering profile
Before a tempering profile is started, disable the screen saver and the energy-saving mode
of your computer, as these could interfere with the programme sequence.
For a tempering profile to be started,
►
it must have previously been saved (see page 36 )
►
the relevant oven must be connected and logged on to the computer and be in operating
mode "I". If this is not the case, an error message will be displayed. In this case, you should
check that the oven is switched on, connected and logged in to CELSIUS.
Every tempering profile can only be run on the oven type for which it was created. The same
applies for saving to a MEMory Card (XL).
A tempering profile can be started
►
via the
button in the toolbar
►
via "Run"→"Start" in the menu bar