beautypg.com

2 mobile heat extractor – application, 3 cold extractor – application, 2 mobile heat extractor – Leica Biosystems CM1520 User Manual

Page 89: Application, 3 cold extractor

background image

89

Leica CM1520 – Cryostat

10.2 Mobile heat extractor –

application

Specimen freezing with the freezing shelf can be accelerated by the ad-

ditional use of a heat extractor.
• Store the heat extractor in the cryochamber.
• Place it on the specimen surface to accelerate freezing.
• Remove it once the specimen is entirely frozen.

10. Ordering Information, Components and Consumables

Fig. 118

7

6

Fig. 119

• Place the cap (9) on the required side, so that

the appropriate location hole for the specimen

disc is visible.

• Fit the pin (6) of the specimen disc (5) into the

corresponding bore (7) on the cold extractor.

• After about 20 seconds, the frozen specimen

can be removed from the specimen disc with

forceps.

• If the cap is too loose, readjust it with the small

screw (10).

Do not overtighten the screw.
• After removing the specimen, remove the cold

extractor from the cold cryochamber once

again.

9

8

5

10

Position the speci-

men here using

forceps

Keep the thermal block outside the cryochamber at room tempera-

ture. Place it in the cryochamber only for specimen removal.

10.3 Cold extractor –

application

The thermal block (8) facilitates the removal of the frozen specimen from

the specimen disc.

Recommendation:

It is recommended to precool the heat extractor in liquid nitrogen or

other refrigerant.