5 manual defrosting of the cryochamber, 6 setting the section thickness, Operating the instrument – Leica Biosystems CM1520 User Manual
Page 29
29
Leica CM1520 – Cryostat
6. Operating the Instrument
6.2.5 Manual defrosting of the cryochamber
You can activate manual defrosting of the cryochamber by pressing the
key (continuous tone sounds), followed immediately by the
key in
the cryochamber temperature selection field (continuous tone stops). The
display flashes during the defrosting (duration 12 min.).
To shut off the manual defrost early, press the
key, followed immediately
by the
button in the cryochamber temperature selection field.
Fig. 19
6.2.6 Setting the section thickness
Fig. 20
Adjust the section thickness using the
and
pushbuttons in the top
control panel field (see Fig. 20) in a range from 2 - 60 µm:
From 2 - 5 µm in 0.5 µm increments
From 5 - 20 µm in 1 µm increments
From 20 - 60 µm in 5 µm increments
The preselected section thickness appears in the display above the keys.
• Start trimming at approx. 20 µm.
• Decrease the section thickness continually down to the appropriate value.
After changing the section thicknesses, discard the first two or three
sections.
• For sectioning, turn the handwheel at a consistent speed.