Leica CM1520 – Cryostat

6. Operating the Instrument

6.2.5 Manual defrosting of the cryochamber

You can activate manual defrosting of the cryochamber by pressing the

key (continuous tone sounds), followed immediately by the

key in

the cryochamber temperature selection field (continuous tone stops). The

display flashes during the defrosting (duration 12 min.).
To shut off the manual defrost early, press the

key, followed immediately

by the

button in the cryochamber temperature selection field.

Fig. 19

6.2.6 Setting the section thickness

Fig. 20

Adjust the section thickness using the

and

pushbuttons in the top

control panel field (see Fig. 20) in a range from 2 - 60 µm:

From 2 - 5 µm in 0.5 µm increments

From 5 - 20 µm in 1 µm increments

From 20 - 60 µm in 5 µm increments

The preselected section thickness appears in the display above the keys.
• Start trimming at approx. 20 µm.
• Decrease the section thickness continually down to the appropriate value.

After changing the section thicknesses, discard the first two or three

sections.

• For sectioning, turn the handwheel at a consistent speed.