8.2 Temperature control button

Possible causes and remedies:
• Temperature of the direct surroundings is

constantly higher than 40 °C.

-->Drop the temperature of the direct surround-

ings.

• When setting up the instrument the minimum

distance (

see "Transport and installation")

to

walls and furniture was not kept.

--> Keep the minimum distance.

• The ventilation slits of the liquefier are dirty.

--> Clean air inlet openings (

see Fig. 54

).

Fig. 51

1

After eliminating the possible sources of error, push the temperature control button (1,

Fig. 51

)

to turn the instrument back on and make it ready to operate. If the instrument does not respond,

contact service.

On the back of the cryostat cabinet there is a temperature control button (1,

Fig. 51

). If the

temperature of the cryochamber exceeds 60 °C the switch is automatically activated and turns

the instrument off.

8. Troubleshooting