9.3.2 Replacing the fuses
9.
Cleaning, Disinfection, Maintenance
On the back of the instrument there is a fuse box with 4 fuses:
• Unscrew the fuse cap with a screwdriver.
• Remove both fuse cap and fuse.
• Insert the defined fuse into the sleeve, then use the screw-
driver to screw the sleeve back in as far as it will go.
Fuse Protection
Type
F1
Display
T 0.25 A
F2
Coarse feed
T 1.6 A
F3
Processor board supply
T 1.0 A
F4
Peltier element
T 6.25 A
Fig. 56
Turn the instrument off and disconnect the power plug before replacing the fuses.
Use only the fuse types specified in Chapter "
3. Technical Data
".
Failure to comply with this can cause severe damage to the instrument and its surroundings
and a risk of fatal injury!
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)