beautypg.com

2 stationary heat extractor (optional), 2 specimen discs, 2 specimen orientation – Leica Biosystems CM1520 User Manual

Page 32: 2 stationary heat extractor, Optional), Specimen discs, Daily use of the instrument

background image

32

Instructions for Use V1.3 – 10/2012

7.

Daily Use of the Instrument

7.2.1 Inserting the specimen discs into the

specimen head

7.2 Specimen discs

Fig. 24

• Lock the handle of the handwheel in the upper

position.

• If the knife holder and a knife are in place,

cover the knife edge with the knife guard.

• Loosen the clamping screw (1,

Fig. 24

) on the

specimen head.

• Insert the shaft of the specimen disc (3) in the

location hole (2) of the specimen head.

• Retighten clamping screw (1).

7.2.2 Specimen orientation

• Loosen the clamping lever (4).
• Orient the specimen on the ball shell using the

lever (5).

• Retighten the clamping lever (4).

7.1.2 Stationary heat extractor

(optional)

• Fasten the support (1,

Fig. 23

) of the heat extractor (2,

Fig. 23

) to the left

side panel of the cryochamber using two screws in the provided holes

and insert the heat extractor.

• Lower the heat extractor cylinder on the specimen surface. After at least

30 seconds, the specimen is completely frozen.

• After the specimen is frozen, move the heat extractor into resting position

at the parking station (3,

Fig. 23

).

1

3

2

4

5

The specimen discs are to be handled

by their side O-rings to protect against

frostbite!

Make sure that the shaft of the speci-

men disc is fully inserted. The entire

surface of the rear side of the speci-

men disc must be free of residue.