Technical data – Leica Biosystems CM1520 User Manual
Page 14

14
Instructions for Use V1.3 – 10/2012
3.
Technical Data
Type
CM1520 CM1520 CM1520 CM1520 CM1520
test
symbol - c_CSA_us
- - -
Nominal voltage
100 VAC
120-127 VAC
220-230 VAC
220-230 VAC
240 VAC
±10 %
±10 %
±10 %
±10 %
±10 %
Nominal frequency
50/60 Hz
60 Hz
60 Hz
50 Hz
50 Hz
Power input
1440 VA
1440 VA
1440 VA
1440 VA
1440 VA
Max. start-up current for 5 sec 45 A eff.
45 A eff.
45 A eff.
45 A eff.
45 A eff.
Protective
class
I I I I I
Pollution degree
2 2 2 2 2
Overvoltage
inst.
category
II II II II II
Heat emission (max.)
1600 J/s
1600 J/s
1600 J/s
1600 J/s
1600 J/s
Refrigeration system
50 Hz/60Hz
Cryochamber
Temperature range
0 °C to -30 °C (+ 3 K/- 3 K)
Cooling time down to -30 °C
max. 6 hrs., at 22 °C ambient temperature
Defrost
Automatic hot gas defrost
1 automatic defrost cycle/24 hours, time-controlled (duration 12 min.)
Refrigeration capacity
690 W
Safety factor
3
Refrigerant
300g ±5g refrigerant R 404A *
Compressor oil
0.6 l EMKARATE RL-22S, ICI *
Quick-freeze shelf
Maximum temperature
- 35 °C (+ 3 K/- 5 K)
Number of freezing stations
10
Defrost
Manual hot gas defrost, time-controlled (duration 12 min.)
according to IEC-1010, UL 3101
according to CECOMAF Liquid temperature 45 °C, evaporation temperature -25 °C
Operating temperature range (ambient temperature): 18 °C to 35 °C. All specifications related to
temperature of the cooling unit are valid only for an ambient temperature of 22 °C and a relative
humidity of no more than 60 %!