Leica CM1520 – Cryostat

53

8. Troubleshooting

8.3 Possible sources of error, causes and remedies

Problem

Cause

Remedy

Frost on chamber walls and micro-

tome.

- Cryostat is exposed to air cur-

rents (open windows and doors,

air conditioning).

- Sliding window was open and

exposed to air currents too long.

- Frost built up by breathing into

the cryochamber.

- Change place of installation

for the cryostat.

- Close the sliding window

correctly.

- Wear a mask if necessary.

Ice formation on the bottom of the

cryochamber

- Drain of the quick freeze shelf

defrosting water obstructed.

- Level the instrument.

Lubricate sections.

- Specimen not cold enough.

- Knife and/or anti-roll plate not yet

cold enough - the section melts

as a result.

- Select lower temperature.

- Wait until knife and/or anti-

roll plate have reached cryo-

chamber temperature.

Sections splinter, cracks in the

section,

- Specimen too cold.

- Select higher temperature.

Sections not properly flattened.

- Static electricity/air currents.

- Specimen not cold enough.

- Large area specimen.

- Anti-roll plate poorly positioned.

- Anti-roll plate poorly aligned with

knife edge.

- Incorrect clearance angle.

- Knife/blade blunt or damaged.

- Eliminate cause.

- Select lower temperature.

- Trim the specimen parallel;

increase section thickness.

- Reposition anti-roll plate.

- Align correctly.

- Set the correct clearance

angle.

- Use different part of the cut-

ting edge or replace.