7. Daily Use of the Instrument

Clearance angle adjustment

Adjust the clearance angle with every new tis-

sue type!

1. The clearance angle scale is located on the

left side of the blade holder.

2. Release the blade holder by turning the No. 4

Allen screw (23) counterclockwise. Select

a clearance angle of 0°. To do so, align the

number 0 to the index mark (25) and tighten the

Allen screw (23). If the sectioning results are

not satisfactory, increase the clearance angle

in 1° increments until you achieve optimum

results.

25

Fig. 35

23

• A clearance angle that is too steep or too shallow may not lead to optimum sectioning results

and can possibly damage the specimen. We generally recommend using higher clearance

angles for harder specimens and relatively lower clearance angles for softer ones.

Settings of 1° - 2° (blade holder CE) are

well suited for most applications.

The harder the specimen is, the larger the clearance angle must be.

However: The larger the clearance angle is, the larger the section buckling becomes.
If the clearance angle is too small, sections that are too thick or thin can result. Also

consider that the position of the knife edge relative to the specimen can change when

the clearance angle is adjusted. For this reason, always place the specimen above the

knife when adjusting the clearance angle. Otherwise, the specimen could hit the knife

while being raised.