beautypg.com

Installation of electronic module, Draining – JULABO CF41 Cryo-Compact Circulators User Manual

Page 75

background image

Cryo-Compact Circulators

75

13. Installation of electronic module

Caution:

Always turn off the unit and disconnect the mains cable from the power source before
performing any service or maintenance procedures, or before moving the unit.
Configuration, installation, maintenance and repairs on the Cryo-Compact Circulator
may only be carried out by qualified personnel.

ALARM

REG+
E-PROG

STAND-BY

A

B

 Turn off the unit with the mains switch and disconnect the

power plug.

 Open the screws (A) and remove the dummy plate.
 Carefully insert the upper edge (B) into the rear panel from

below, align the electronic module at top and bottom and
applying light pressure connect the 15-pole connector.

 Attach the electronic module using the screws (A).
 The Cryo-Compact Circulator is ready for use. It automatically

recognizes the installed electronic module.



13.1. Draining

Notice:

Do not drain the bath fluid while it is hot!

Check the temperature of the bath fluid prior to draining (by switching the unit on for
a short moment, for example).

Store and dispose the used bath fluid according to the laws for environmental
protection.


Draining

 Turn off the unit and disconnect the mains cable from the

power source.

 Place the Cryo-Compact Circulator near the rim of the table.

Use a suitabel vessel as recipient for the bath liquid

 Unscrew the drain tap and empty the unit completely.
 Tighten the drain tap.