beautypg.com

Control, S xx.xx e xx.xx – JULABO CF41 Cryo-Compact Circulators User Manual

Page 36

background image

Menu functions

36

8.2. Control

The Cryo-Compact Circulator is conceived for internal and external
temperature control. Switching is carried out in this submenu. Depending on
what is set, only the respective set of parameters is indicated.

Main menu

Profile Start
Int. Programmer

>Control

Control
>ControlType:

Selftuning : once /always/off

Dynamic : aperio /normal
Xp : 1.0K
Tn : 100s
Tv : 5s

intern

Control
>ControlType: extern

Selftuning: once /always/off
CoSpeed : 0.0
Xp : 0.7K
Tn : 720s
Tv : 55s
Xpu : 5.0K

1

ESC

ESC

internal

external

Example: 0.8

1.

Use the keys

to select the menu item.

2.

With the keys

select the parameter (parameter blinks)

or
use the numeral keypad to enter the numbers (number segment blinks).

3.

Press enter

to store the new parameter or value.

-OFF-

S xx.xx
E xx.xx

ControlType:

Can only be adjusted in the -OFF- condition

Possible parameters:

intern

internal temperature control

extern

external temperature control with external Pt100 sensor

Pt100

IMPORTANT:
Additional measures for external temperature control.

For external control and temperature measurement an external Pt100
sensor must be connected to the socket (7) on the rear of the Cryo-
Compact Circulator .