beautypg.com

12 3 ak – JULABO CF41 Cryo-Compact Circulators User Manual

Page 67

background image

Cryo-Compact Circulators

67

4

1

3

2

ext. Pt100

Socket for external Pt100 sensor

1

4

2

3

Shield

Plug

Look on

soldering side.

Pt100



Pin assignment:
Pin

Signal

1

I+

2

U+

3

U-

4

I-

The shield of the connecting cable is electrically connected to the plug
housing and the sensor tube.

ALARM

Alarm output

(for external alarm signal)
This potential-free change-over contact is activated in case of an alarm
when pins 2 and 3 are connected.

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

/

Switching current

max. 1 A

1

2

3

AK

STAND-BY

STAND-BY input

(for external emergency switch-off)

Pin assignment:

Pin

Signal

1

not

connected

2

5 V / DC

3

0 V


Functional description see page 54

1

2

3

4

5

6

REG+E-PROG

Programmer input / temperature recorder output

Pin

Signal

1 Voltage output

Channel 1

0 ... 10 V

2 Voltage output

Channel 2

0 ... 10 V

3 GND for outputs

0 V

4 Programmer input

EPROG

0 to 10 V / 0 to 20 mA

5 Current output

Channel 3

0 to 20 mA / 4 to 20 mA

6 GND for Programmer

0 V

Functional description see page 51

This manual is related to the following products: