Menu functions

60

Type:

A >1-point<, >2-point < or >3-point < calibration can be carried
out.
First geometrically define the location for calibration (measuring point
T

M

), then determine the temperature values of of the calibration

points.
The calibration type also determines the number of pairs of values
indicated on the LCD DIALOG-DISPLAY.

TmpValue:

1 or 2 or 3

Defined temperature value of the calibration point. This value is
automatically stored with >CalValue< and can be indicated for control
purposes.

CalValue:

1 or 2 or 3

The „Calibration value“ is determined with a temperature measuring
device and stored under menu item >CalValue<.

Example:
1-point calibration

°C

°C

T 1

T


3-point calibration

°C

°C

T 1

T

T

T

1 = Original curve

Principle:
For ATC calibration, in steady state the bath temperature at the location
of the temperature sensor (T

M

) is determined at the respective adjusted

working temperature. This value is then set on the Cryo-Compact
Circulator in the menu >ATCalibration< under menu item >Calvalue X<.
This can be a 1-point, 2-point or 3-point calibration.

°C

T

T

M

B (EXT)

Pt100

ext

T

T

M

M = Temperature measuring instrument with temperature sensor
B = Bath tank (internal or external)
T = Cryo-Compact Circulator

T

M

= Temperature on measuring point

T

T

= Temperature on Cryo-Compact Circulator