JULABO CF41 Cryo-Compact Circulators User Manual

Page 59

background image

Cryo-Compact Circulators

59

8.8. ATC Absolute Temperature Calibration, 3-point calibration

ATC serves to compensate a temperature difference that might occur between
Cryo-Compact Circulator and a defined measuring point in the bath tank
because of physical properties.

Main menu
Interface serial

Pump

>ATCalibration

ESC

ESC

ESC

1

1

ATCalibration
>Sensor : intern

Status : no /yes
Type : 3-point /1-point/2-point
TmpValue1: xx.xx°C
CalValue1: xx.xx°C
TmpValue2: xx.xx°C
CalValue2: xx.xx°C
TmpValue3: xx.xx°C
CalValue3: xx.xx°C

ATCalibration
>Sensor : extern

Status : no /yes
Type : 3-point /1-point/2-point
TmpValue1: xx.xx°C
CalValue1: xx.xx°C
TmpValue2: xx.xx°C
CalValue2: xx.xx°C
TmpValue3: xx.xx°C
CalValue3: xx.xx°C

internal

external

Sensor: >intern<, >extern<

Calibration can be carried out for the internal temperature sensor and
for the external temperature sensor connected to the socket „ext.
Pt100“.
The Cryo-Compact Circulator is able to store both parameter sets.
However, only the one set under this menu item is indicated.

Status:

>no<

The controller of the Cryo-Compact Circulator uses the

original curve of the temperature sensor.
Important: During the calibration process >no< needs to be set.

>yes<

The controller of the Cryo-Compact Circulator uses the

new calibration curve.