37

Sensor calibration of the external Pt100 sensor is carried out in the menu
>ATCalibration<, with the > Status< set to >no< (see page 59).

Suggested adjustments for external temperature control:
AreaUp/ AreaDown and IntMax / IntMin
see chapter >Limits< on page 56.

External temperature control does not allow for setpoint setting via the
socket „ext. Pt100“ (see page 32).

Pt100 M+R

Accessory: Pt100 external sensor

Order No.

Description

Material

Cable

8981003

200x6 mm Ø,

stainless steel

1.5 m

8981005

200x6 mm Ø,

glass

1.5 m

8981006

20x2 mm Ø,

stainless steel

1.5 m

8981010

300x6 mm Ø,

stainless steel

1.5 m

8981015

300x6 mm Ø,

stainless steel / PTFE coated

3 m

8981013

600x6 mm Ø,

stainless steel / PTFE coated

3 m

8981016

900x6 mm Ø,

stainless steel / PTFE coated

3 m

8981014

1200x6 mm Ø,

stainless steel / PTFE coated

3 m

8981103 Extension cable for Pt100 sensor

3.5 m

8981020 M+R in-line Pt100 sensor

The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit

Notice:

Place the external sensor into the bath medium and securely fix the sensor.



Selftuning:

When performing an selftuning for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv are automatically
determined and stored.

Possible parameters:

off - no selftuning.

The control parameters ascertained during the last identification are used

for control purposes.

once – single selftuning (factory setting)

The instrument performs a single selftuning of the controlled system after

each start with the start/stop key

or after receiving a start command

via the interface.

always – continual selftuning

The instrument performs a selftuning of the controlled system whenever a

new setpoint is to be reached.

Use this setting only when the temperature application system changes

permanently.