Operating controls and functional elements

16

1

Mains power switch, illuminated

2

°C

140

100

180

220

0

60

20

Adjustable excess temperature protection according to
IEC 61010-2-010

3

Setpoint1: 150.00°C
IntAct : 148.53°C
Power : 80 %
Control : intern

RS232 : 150.00°C
ExtAct : 145.30°C
Power : 80 %
Control : extern

LCD DIALOG-DISPLAY

Line 1: Setpoint and origin of setpoint programming

(Key / RS232 or RS485 / ext. Pt100 / EProg)

Line 2: Actual value - internal or external,

identical to line 1 of the VFD-COMFORT-DISPLAY

Line 3: Heating capacity in %.

Line 4: Control type: internal / external

4

VFD-COMFORT-DISPLAY

Header: Control indicators see sections 4.1 and 4.2

Line 1: Actual value internal or external

The display is depending on the selected control mode in the
menu > Control < (internal or external).

Line 2: Working temp. setpoint, constantly S xxx.xx

Line 3: Actual value (E = external or I = internal)

Alternating with the display in line 1

Use the keys

to indicate further values in line 3. However,

the functions of these keys are different with the programmer
started.
PI

Capacity in % - with manipulated variable set to >control<*

PS Capacity in % - with manipulated variable set to >serial<* or
>eprog<*
H

Heater capacity in Watts

U

Mains voltage Volts

F

Flow rate in liters/minute

(providing EPROG input set to >Flowrate<)

*see >Configuration> page 32

4.1

Control indicators in the header:

Heating / Cooling / Alarm /

R

emote control

4.2

Control indicators in the header:
Temperature indication Internal or External actual value
Temperature indication in °C or °F (see page 35)