JULABO CF41 Cryo-Compact Circulators User Manual
Page 64
Troubleshooting guide / Error messages
64
ALARM
CODE 06
Difference between
working- and safety-
temp. Bigger than
permissible value.
Defect of the working or excess temperature sensor.
Working temperature and excess temperature sensors report a
temperature difference of more than 35 K.
ALARM
CODE 07
The internal I2C
bus-system shows
a failure.
Other errors (I
2
C-BUS errors)
ALARM
CODE 12
Error in A/D-
converter of the
internal measuring
system.
Error in A/D converter
ALARM
CODE 14
Working temperature
exceeds the adjusted
safety temperature.
Excess temperature sensors defect.
The excess temperature value lies below the working temperature
setpoint. Set the excess temperature to a higher value.
ALARM
CODE 15
External control
selected, but ext.
sensor not connected
or defective.
External control selected, but external Pt100 sensor not connected
or defect.
WARNING
CODE 20
Cooling is affected
or ambient temperature
to high
Clean condenser
Cooling of the condenser is affected. Clean air-cooled condenser.
Check the flow rate and cooling water temperature on water-cooled
condenser.
WARNING
CODE 21
Compressor does not
work. Compressor
will be automatic-
ally reconnected.
Compressor does not work.
After a short cooling interval, the compressor motor will be
automatically reconnected and the message "CODE 21" no longer
appears.
ALARM
CODE 33
Internal safety-
temperature sensor
short-circuited or
interrupted.
Cable of the excess temperature sensor interrupted or short-circuited