beautypg.com

S 25.00 e, S 58.00 e – JULABO CF41 Cryo-Compact Circulators User Manual

Page 41

background image

Cryo-Compact Circulators

41







Examples:

-OFF-

S 25.00
E -----

Setpoint3: 80.00°C
IntAct : 80.00°C
Power : 2%
Control : intern


Setpoint
Setpoint1: 25.00°C
>Setpoint2: 58.00°C
Setpoint3: 85.00°C

Setpoint2: 58.00°C
IntAct : 60.00°C
Power : 0%
Control : intern

60.00

S 58.00
E -----

Profile End Stdby / SetpPG / Setp 1 / Setp 2 / Setp 3

This parameter is used to determine if the unit switches to the -OFF-
condition at the end of a profile, or if temperature control is continued
and the working temperature setpoint that is be used in such case.

Stdby – If the end of the profile is reached, the Cryo-Compact
Circulator switches to the -OFF- condition. The VFD display
indicates the working temperature setpoint, that was displayed in line
2 already before the program was started (example: S 25.00 °C ).

SetpPG – (Setpoint of the ProGrammer) At the end of the profile, the
setpoint of the last program section is written to the position of
setpoint 3 (example: 80.00 °C) and indicated in line 1 of the LCD
display. The Cryo-Compact Circulator operates and controls this

temperature until the Start/Stop key

is pressed or a new

profile is started.

Setp 1, Setp 2, Setp 3 – („Setp“ short for Setpoint)
Before starting the programmer, a temperature value is set for one of

the three setpoints in the

menu.

(example: Setpoint 2 58.00 °C)
The Cryo-Compact Circulator operates and controls this

temperature at the end of the profile until the Start/Stop key

is

pressed or a new profile is started.

Start

no / yes

 (manual start)

or

time

 (via integrated timer)

When selecting the parameter time, a new submenu is called up for
entry of the start time.
A flashing segment indicates that a start time needs to be entered.
hour.min

Start time

Day.Mon

day and month

Year

year

Set each entry with enter

.


Start no / yes
A flashing line indicates that the parameter „yes“ needs to be entered.

Press the key

to select the parameter and press enter

.

Starttime: 06:00
Date : 15:12
*** wait ***
Acttime 02:34:45

The temperature system switches to waiting mode and a flashing line
„wait“ appears on the LCD DIALOG-DISPLAY. The start time and actual
time are permanently indicated on the display.

This manual is related to the following products: