JULABO CF41 Cryo-Compact Circulators User Manual
Page 29
Cryo-Compact Circulators
29
OverTemp 87.0°C
SubTemp 83.0°C
Setpoint3 85.0°C
The warning functions are only triggered when the actual bath temperature,
after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3
seconds.
Setting range:
20 °C ... 220 °C
SafeTemperature: Indicated is the cut-out value set with a screwdriver on the
excess temperature protection device.
This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a complete
shutdown of the heater and pump is effected.
The alarm is indicated by optical and audible signals (continuous tone).
The following error message appears on the VFD COMFORT-DISPLAY:
XXXXX
ALARM
CODE 14
Press
for help on the LCD
Working temperature
exceeds the adjusted
safety temperature.
SafeSens: Indicated is the temperature value of the safety sensor.