beautypg.com

JULABO CF41 Cryo-Compact Circulators User Manual

Page 43

background image

Cryo-Compact Circulators

43







>Edit Profile: 1
**Step is deleted **
Setp: °C
Time [hh:mm]

Edit


>Edit Profile: 1
** Step is active **
Setp: 100.00°C
Time [hh:mm] 01:05

Edit

Deleting a complete section:
 Use the keys

to move to line 2.

 Use the keys to choose the desired section.

 Press

the

key to delete the complete section.

If the desired section is already chosen, the section is deleted even if the

cursor > is in line 1, 3 or 4.
Deletion is indicated in the respective line of the LCD display:
** Step is deleted **

The section that is currently processed cannot be deleted. If you still
try it, the message ** Step is active ** appears on the
LCD display.

Edit
Step: 6
>Setp: 83.00°C
time [hh:mm] 01:25

Changing a section:

The values in lines 3 and 4 can be changed, even on the section that
is currently processed.

 In the input window, select the menu item with the keys

.

 Use the numeral keypad to enter the numbers and press enter

.


Examples:

Setpoint:

target temperature 83 °C

Time

time period 1 hour:25 minutes

Particularity for changes in a section that is currently in process.

If the target temperature is changed, the slope of the ramp is
automatically calculated on basis of the remaining time of the
section.

If the time period is changed, the slope of the ramp is automatically
calculated on basis of the new total time of the section.