Operating procedures, Power connection, Filling – JULABO CF41 Cryo-Compact Circulators User Manual
Page 23
Cryo-Compact Circulators
23
6. Operating
procedures
6.1. Power
connection
Caution:
Only connect the unit to a power socket with an earthing contact (PE – protective
earth)!
The power supply plug serves as a safe disconnecting device from the line and
must always be easily accessible.
Never operate the unit with a damaged mains power cable.
Regularly check the mains power cables for damage.
We disclaim all liability for damage caused by incorrect line voltages!
Make sure that the line voltage and frequency match the supply voltage
specified on the type plate.
6.2. Filling
Notice:
Pay attention to the thermal expansion of bath oil during heating to avoid overflowing of
the liquid.
Guideline:
A volume change of 12 % per 100 °C temperature variation is to be considered.
Take care that no liquid enters the interior of the Cryo-Compact
Circulator .
Connect the tubing from the external system to the pump
connectors and check for leaks
Check to make sure that the drain tap (7) is closed.
Recommendation:
For filling, use for example an measuring jug with nuzzle.
Recommended maximum filling level with water as bath fluid:
30 mm below the tank rim
Recommended maximum filling level with bath oils:
40 mm below the tank rim
Turn the mains switch (1) on
(Switching on - see page 24)
Switch on unit by pressing the start/stop key
.
Tempering fluid is pumped into the externally connected system.
Refill fluid up to recommended filling level.
The Cryo-Compact Circulator is ready for operation.