8.6. Limits

The menu >Limits< allows for setting the minimum and maximum values for all
important setting ranges and capacity variables.

Main menu

Interface serial
ATCalibration

>Limits

ESC

ESC

1

Limits
>SetMax :300.00°C
SetMin :-94.99°C
HeatingMax: 100%
CoolingMax: 0%
IntMax :300.00°C
IntMin :-94.99°C
Upper band: 200K
Lower band: 200K

In case of external control
these menu items are
additionally indicated

SetMax: and SetMin:

Maximum and minimum setpoint

Restriction for the adjustable temperature range.

Setting range: see working temperature range

The limitation of the operating temperature range effects the temperature

setting under the menu called up with the

key - Setpoint 1, 2, 3.

It is possible to adjust only working temperatures that lie within the limit range
set here.
Existing settings for Setpoint 1, 2, 3 and also for >OverTemp< and >SubTemp<
(see page 27) are automatically defered within the limit range.

HeatingMax: and CoolingMax:

Adjusted maximum heating / cooling.

Setting range: 0 to 100 % in steps of 1 %

Heating and cooling powers of the Cryo-Compact Circulator are adjustable.
100 % corresponds to the values in the technical specifications of the
equipment.

Factory setting: CoolingMax

CF31 100 % (fixed value)

CF41 0 to 100 % in steps of 1 %