Menu functions

32

8.1. Configuration

>Configuration
Control
Profile Start

Main menu

Configuration
>Setpoint : keypad /RS232/ext.Pt100/eprog
Autostart: off /on
OFF-Mode : pump off /pump on
Language : German /Englich
Reset : no /yes
Variable : control /serial/eprog
Unit : °C /°F

ESC

ESC

1.

Use the up/down keys

to select the desired menu item.

2.

Use the left/right keys

to select the parameter (parameter blinks).

3.

Press enter

to store the new parameter.

Setpoint1: 150.00°C
IntAct : 148.53°C
Power : 80 %
Control : intern

RS232 : 150.00°C
ExtAct : 145.30°C
Power : 80 %
Control : extern

-OFF-

S xxxx
I xxxx

R

Pt100 : 150.00°C
ExtAct : 145.30°C
Power : 80 %
Control : extern

Setpoint

The Cryo-Compact Circulator provides four possibilities for setpoint setting. The
selected mode is indicated in line 1 on the LCD DIALOG-DISPLAY.

Possible parameters:

keypad – Setpoint setting via keypad or via the integrated programmer.

(factory setting)

RS232 or RS485 - Setpoint setting via the serial RS232/RS485 interface

through a PC or superordinated data system.

In the header of the VFD COMFORT-DISPLAY an „

R

“ illuminates.

It indicates that remote control mode is set.

RS232 or RS485: see menu > Serial interface < page 58.

Important:

Connect the Cryo-Compact Circulator to a PC using an interface cable.
Check the interface parameters of both interfaces (on Cryo-Compact
Circulator and PC) and make sure they match.
(see 11.1.

on page 68 )

ext.Pt100 – Setpoint setting via the analog socket „ext. Pt100“ using an

external temperature sensor or an appropriate voltage/current source.