Safety installations, warning functions – JULABO CF41 Cryo-Compact Circulators User Manual
Page 27
![background image](/manuals/563121/27/background.png)
Cryo-Compact Circulators
27
7.4.
Safety installations, warning functions
Check the safety installations at least twice a year! (See page 14)
7.4.1.
Excess temperature protection
Settings for the excess temperature protection according to IEC 61010-2-010
and for the high and low temperature warning functions are made
in a menu that is called up with the key
.
ESC
1
Safety-values
>Warn-Type: Warning /Alarm
OverTemp : 85.00°C
SubTemp : -99.99°C
SafeTemp : 50.00°C
SafeSens : xx.xx°C
°C
°C
100
140
220
0
60
20
Legend:
Setting via keypad
Select parameter
Select menu item
Change menu level
scroll down / scroll up
1
Example:
1.
Press the key
The LCD DIALOG-DISPLAY shows the valid settings (see example above).
2.
Use the keys
to select the menu item.
Settings:
Warn-Type: Select the parameter with the key
and press
enter
(warning or alarm).
OverTemp and SubTemp: Use the numeral keypad to enter the value and
press enter
.
SafeTemp: Set the new cut-out value using a screwdriver via the LCD
DIALOG-DISPLAY. (Example: 50 °C)
The value remains visible on the LCD DIALOG-DISPLAY for about 10
seconds, or press
to update the display immediately.