Safety installations, warning functions

Check the safety installations at least twice a year! (See page 14)

7.4.1.

Excess temperature protection

Settings for the excess temperature protection according to IEC 61010-2-010
and for the high and low temperature warning functions are made

in a menu that is called up with the key

.

ESC

1

Safety-values
>Warn-Type: Warning /Alarm
OverTemp : 85.00°C
SubTemp : -99.99°C
SafeTemp : 50.00°C
SafeSens : xx.xx°C

°C

°C

100

140

220

0

60

20

Legend:

Setting via keypad

Select parameter

Select menu item

Change menu level
scroll down / scroll up

1

Example:

1.

Press the key

The LCD DIALOG-DISPLAY shows the valid settings (see example above).

2.

Use the keys

to select the menu item.

Settings:

Warn-Type: Select the parameter with the key

and press

enter

(warning or alarm).


OverTemp and SubTemp: Use the numeral keypad to enter the value and

press enter

.


SafeTemp: Set the new cut-out value using a screwdriver via the LCD
DIALOG-DISPLAY. (Example: 50 °C)

The value remains visible on the LCD DIALOG-DISPLAY for about 10

seconds, or press

to update the display immediately.