JULABO CF41 Cryo-Compact Circulators User Manual
Page 42
![background image](/manuals/563121/42/background.png)
Menu functions
42
Examples:
Setp. : 125.8°C
IntAct: 124.7°C
Prof. : 4 Step: 12
Remain: 00:03:45
Setp. : 125.8°C
IntAct: 124.7°C
Prof. : 4 Step: 12
All : 00:03:45
Setpoint : 125.8°C
IntAct : 124.7°C
Prof : 4 Step:12
S12:150:00 h:m 01:10
>Edit Profile: 4
Step: 12
Setp: 150.00°C
Time [hh:mm] 01:10
Edit
Indication after successful start:
The started programmer displays the actually calculated setpoint in line
1. The value changes until the final temperature of the section is reached
(temperature ramp).
1
st
line:
Setpoint of the programmer
2
nd
line:
Actual temperature value
at
internal control = IntAct:
xxx.xx
at
external control = ExtAct: xxx.xx
3
rd
line:
Selected profile and the actual section (step).
4
th
line:
With the keys
the following information can be
displayed in line 4.
Remaining time of the actual section (standard display)
Remaining total time: Profile x number of repetitions
(All: Days/hours:minutes:seconds)
Particularitys:
1. If a total time of more than 999 days is calculated, only the hours are
displayed on the unit.
(All: hours:minutes:seconds)
2. If the time period of a section is edited after start, then the remaining
total time can no longer be displayed.
The data of a section as stored under „Edit Profile“.
Use the keys
to successively indicate the sections of a profile.
Unused sections or incorrect data are skipped.
8.3.1.
Edit after start
A JULABO programmer of generation II allows editing of a running
profile. A programmer of the previous generation requested the profile to
be interrupted, edited and then started again.
>Edit Profile: 1
Step: 6
Setp: 100.00°C
Time [hh:mm] 01:05
Edit
Start and quit the edit mode:
After start, press the
key to enter the submenu „Edit Profile“.
Press
to quit the edit mode.