Cryo-Compact Circulators

55

8.5.4.

ALARM output


ALARM


Alarm output (9)

(for external alarm signal)
This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs< all operating conditions
can be signaled without having to change the pin assignments.

Menu >Inputs/Outputs<

AlarmOut
>Function: StdBy /Alarm/Al+StBy
Type : normal /invers

Signification of the terms under menu item >Function< :

The Cryo-Compact Circulator is in condition
>StandBy< or

>Alarm<

-OFF-

S xx.xx
E -----

XXXXX

ALARM
CODE xx

For >Type: normal< pins 2 and 3 are connected in any case
according to the selected >Function<.

For >Type: invers< pins 2 and 1 are connected in any case
according to the selected >Function<.

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

/

Switching current

max. 1 A