JULABO CF41 Cryo-Compact Circulators User Manual
Page 55
Cryo-Compact Circulators
55
8.5.4.
ALARM output
ALARM
Alarm output (9)
(for external alarm signal)
This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs< all operating conditions
can be signaled without having to change the pin assignments.
Menu >Inputs/Outputs<
AlarmOut
>Function: StdBy /Alarm/Al+StBy
Type : normal /invers
Signification of the terms under menu item >Function< :
The Cryo-Compact Circulator is in condition
>StandBy< or
>Alarm<
-OFF-
S xx.xx
E -----
XXXXX
ALARM
CODE xx
For >Type: normal< pins 2 and 3 are connected in any case
according to the selected >Function<.
For >Type: invers< pins 2 and 1 are connected in any case
according to the selected >Function<.
Switching capacity
max. 30 W / 40 VA
Switching voltage
max. 125 V
/
Switching current
max. 1 A