beautypg.com

JULABO CF41 Cryo-Compact Circulators User Manual

Page 10

background image

Operator responsibility – Safety instructions

10



CF41

Working temperature range

°C

-40 ... +200

Temperature stability

°C

±0.02

Temperature selection

digital

via

keypad

indication on LCD DIALOG-DISPLAY (°C/°F)

remote control via personal computer

indication on monitor

Temperature indication:

VFD COMFORT-DISPLAY

Resolution

°C

0.01

Pump pressure stage

indication on VFD COMFORT-DISPLAY

Temperature control

ICC - Intelligent Cascade Control

Working temperature sensor

Pt 100

Safety temperature sensor

Pt 100

Absolute Temperature Calibration

at internal control

°C

±3

at external control

°C

±10

Heater wattage

(at 230 V)

kW

2,0

Heater wattage

(at 115 V)

kW

1,0

Cooling capacity

°C

20 0 -20 -30

Medium ethanol

kW

0.47 0.4 0.28 0.12

Cooling compressor

1-stage

Refrigerant

R404A

Cooling machine

Air-cooled

Electrical connections: see page 11

Electronically adj. pump capacity

1 ... 4

Flow rate

min ... max. at 0 bar

l/min

22 ... 26

Pressure max.

at 0 liter

bar

0,7

Suction max

at 0 liter

bar

0,4

Bath opening (WxL)

cm

19x3

Bath depth

cm

19

Filling volume

Liter

4.0 ... 5.5

Overall dimensions (WxDxH)

cm

28x46x46

Weight kg

42

Ambient temperature

°C

5 ... 40

Protection class according to IEC 60 529

IP 21

Mains power connection 230 V/50 Hz

V/ Hz

207-253 / 50

Current draw (at 230 V)

A

13

Mains power connection 230 V/60 Hz

V/ Hz

197-242/60

Current draw (at 230 V)

A

13

Mains power connection 115 V/60 Hz

V/ Hz

103-127 / 60

Current draw (at 115 V)

A

18


All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.

This manual is related to the following products: