8.3 GROUNDING CONSIDERATION

Proper grounding and shield termination is mandatory for accurate measurements.

8.3.1

VOLTAGE MEASUREMENT GROUNDING

If the voltage to be measured is returned to earth ground (within the common mode voltage

range) at its source, neither the Input nor the Common lead should be grounded at the voltage

measurement point, since the RQCM will return to earth ground through its power cord. The

shield for the input leads must only be terminated at the Data Acquisition Card connector.
If the voltage to be measured is isolated from earth ground, the shield or its drain wire should be

connected to the common side of the voltage to be measured at the voltage source as well as at

the shield terminal on the RQCM Data Acquisition Card.

8.3.2

TEMPERATURE MEASUREMENT GROUNDING

All three temperature-sensors must be of the isolated or ungrounded type. Sensor lead wire

shields should be terminated at the RQCM Data Acquisition Card connector only. The measured

device should be connected to earth ground. Exposed junction probes should not be used to

measure the temperature in a conductive media like water.