beautypg.com

Control the temperature, Keep the test chamber clean, Keep a constant gas flow – INFICON RQCM - Quartz Crystal Microbalance Research System User Manual

Page 34: For operation in liquids, Degas the sample liquid, Wait for mechanical disturbances to stabilize, Wait for the temperature to stabilize, Control the temperature -10, Keep the test chamber clean -10, Keep a constant gas flow -10

background image

RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

OPERATION

3-10

drift is temperature. Review section 4.1.9 for more information on the crystal’s response to

temperature.

3.9.3.1

Operate at or Near The Crystal’s “Turn-Around-Point”

When possible, perform the experiment at the sensor crystal’s “turn-around-point”. At or near

this temperature, most of the errors due to temperature are negligible. The “turn-around-point” is

where the temperature coefficient of the crystal is zero. That is, there is no change in resonance

frequency due to a change in the temperature of the crystal at the turnaround point. INFICON’s 1

inch crystals are optimized for two operating temperatures namely 90ºC and 25ºC.

3.9.3.2 Control

The

Temperature

If operation at or near the crystal’s zero turn-around-point is not possible, the operating

temperature should be controlled within a degree or a fraction of a degree of its desired value.

How tightly the temperature should be controlled depends on how far the operating temperature is

from the zero turn-around-point, the further away from the turn-around-point, the larger the error.

For example, for 90ºC-cut-crystal, about 1.5 Hz/ºC error is expected if operating at 80ºC.

However, as much as 5 Hz/ºC error is expected when operating at 60ºC.

3.9.3.3

Keep the Test Chamber Clean

Keep the test chamber clean, as any particle attached to the crystal’s sensing surface will result in

a frequency change.

3.9.3.4

Keep a Constant Gas Flow

If a gas is used, keep the flow constant throughout the experiment. The crystal’s frequency is

also sensitive to changes in pressure, which will result, if the flow rate is not constant.

3.9.4

FOR OPERATION IN LIQUIDS

Since the sensor crystal is dampened by the liquid, any error in the measurements will be

amplified. Extra care must be taken to ensure minimal error. The sample liquid should be

prepared carefully. Changes in temperature or the properties of the solvent as well as air bubbles

will affect the sensor signal.

3.9.4.1

Degas The Sample Liquid

The sample liquid should be degassed prior to measurement to avoid the formation of air bubbles

on the surface of the crystal.

3.9.4.2

Wait For Mechanical Disturbances To Stabilize

To minimize random fluctuations cause by vibrations, it is best to immerse the holder, with a

crystal installed, in the sample solution several hours before the experiment is started.

3.9.4.3

Wait For The Temperature To Stabilize

To avoid the formation of air bubbles and reduce temperature related artifacts, the sample liquid

should have approximately the same temperature as the measurement chamber’s working

temperature (±2°C). Wait at least one hour after a holder is immersed in a liquid to allow for the

crystal to come to equilibrium, before performing any accurate measurements.