RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

CRYSTALS, HOLDERS AND FLOW CELL

4-6

always good practice to wait at least 30 minutes before performing any accurate measurements

after the crystal comes in contact with a new medium. This allows the crystal to come to

equilibrium with the medium. If temperature control is not possible or practical, attempts should

be made to measure the temperature of the solution around the crystal during the experiments and

perform temperature compensation.
In short, each RQCM user must determine the effect of temperature on the experiments being

performed and either control the temperature accordingly, or measure the temperature and

compensate for it.