beautypg.com

Crystal measurements verification, In air, Crystal measurements verification -2 – INFICON RQCM - Quartz Crystal Microbalance Research System User Manual

Page 22: In air -2

background image

RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

GETTING STARTED

2-2

IEEE-488 communication option, install the proper cable.
Refer to Section 7.1 to install, setup and run the RQCM software.
Connect the crystal holder, with a crystal installed, to the SMB connector labeled Crystal

by means of the 24-inch SMB coaxial cable.
Observe the AC voltage setting on the rear panel. Make sure it is set for your local line

voltage. Plug one end of the power cord to a power outlet and plug the other end into the

power entry module in the rear of the RQCM.
Refer to Figure 4, and Figure 7 for complete system connections. If your RQCM is

equipped with optional cards, refer to their appropriate section for detail instruction on

installation and operation.
Switch the front panel power switch to on. All of the red communication LEDS on the

front panel will light up for two seconds then some will turn off reflecting the status of

the communication lines. The green System Ready LED will come on and remain on

until the RQCM power is turned off. If the System Ready LED fails to turn on then there

is an internal problem with the RQCM. Please refer to section 10 for troubleshooting.
Start the RQCM computer program. Note that you may have to set the RQCM address

and select the correct COMM port in the Setup Menu in order for the RQCM to

communicate with your PC. Click on the View Status button to bring up the Status

Screen. The Status Screen should indicate a crystal frequency within the specified range

for the type of crystals being used. The frequency should be stable to within a few hertz

and the crystal resistance should be between 5 Ω and 15 Ω for an uncoated polished

crystal in air.
Check the capacitance cancellation by pressing and holding the Reset switch. The green,

Lock LED should light. Keeping the Reset switch pressed, adjust the fine capacitance

trimmer counterclockwise (decreasing the capacitance) by about 5 degrees. The yellow,

Sweep LED should flash. Back the trimmer clockwise to the point where the Sweep

LED just stops flashing. The capacitance cancellation should be checked and readjusted

every time the environment of the crystal and holder is changed. i.e. if the crystal and

holder are moved from air to liquid or liquid to air.
Remove the crystal. The red, Unlock, LED should light. The green, Lock, LED should

go off. The Sweep LED should not flash. If the Sweep LED flashes the capacitance is

under compensated. Reinstall the crystal in the holder and repeat the process until it is

perfectly compensated (Sweep LED not flashing when the crystal is removed). Refer to

Section 3.4 for more details on adjusting the capacitance cancellation.

2.3.1

CRYSTAL MEASUREMENTS VERIFICATION

A quick way to test your RQCM is to verify its measurements of frequency and

resistance in air and in water. The measurement values of frequency and resistance are as

follows. For a thorough test, the crystal can be immersed in a series of viscous

glycerol/water solutions at 20ºC and compare its measurement values against the

predicted results shown in Figure 14 and Figure 15.

2.3.1.1 In

Air

If you are using INFICON 1” Polished, 5 MHz Gold Electrode Crystal, after compensated

for the capacitance, the frequency should be between 4.976 to 5.020 MHz, and the

resistance should be between 5 to 15 ohms. Record the frequency and resistance values.