beautypg.com

Operation guidelines, Allow the rqcm to warm up, Isolate the rqcm from temperature change – INFICON RQCM - Quartz Crystal Microbalance Research System User Manual

Page 33: For operation in air, Operation, Guidelines -9, Allow the rqcm to warm up -9, Isolate the rqcm from temperature change -9, For operation in air -9, Igure

background image

RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

OPERATION

3-9

Figure 5 Typical Voltammogram Plot Obtained Using the RQCM

3.9 OPERATION

GUIDELINES

This section offers some general guidelines to help minimize errors in the measurements due to

the environment of the sensor crystal. It is the user’s discretion as which are best to utilize for a

particular experiment. Keep the following tips in mind when setting up an experiment.

3.9.1

ALLOW THE RQCM TO WARM UP

Allow for about one to two hours of warm-up time for the RQCM’s electronics to reach

temperature equilibrium.

3.9.2

ISOLATE THE RQCM FROM TEMPERATURE CHANGE

Limit temperature changes of the RQCM’s environment as this will cause small changes in the

measured crystal frequency. The most likely source that alters the RQCM temperature is the

room air-conditioning system. Keep the RQCM away from the direct path of air-conditioning

vents.

3.9.3

FOR OPERATION IN AIR

In gas-phase measurements, the single factor that is most likely to induce frequency instability or