RQCM – RESEARCH QUARTZ CRYSTAL MICROBALANCE

CRYSTALS, HOLDERS AND FLOW CELL

4-

10

4.3 CRYSTAL

HOLDERS

Figure 12 shows a INFICON CHC-100 Crystal Holder (without a crystal, the crystal retainer or the

retainer cover). It has a cavity for a 1-inch diameter crystal. Inside the cavity there are two

Pogo® pins providing connections to the crystal’s front and rear electrodes. Note the locations of

the Pogo® pins. These pins are internally connected to the BNC connector (SMB Jack for CHT-

100 and CHK-100 holders) via an internal coaxial cable.

Figure 12 CHC-100 Crystal Holder

4.3.1

HOW TO INSTALL A CRYSTAL IN A INFICON CRYSTAL HOLDER

1. Identify the Front and Rear Sides of the crystal. See Section 4.1.
2. Clean & Dry the Crystal Holder cavity, then insert the Crystal with the Front Side (Sensing

Electrode) exposed. The “Wrap-Around Extended Electrode” MUST be in the 60º region as in

Figure 13 below.