once a month and recorded to verify if any changes in the quality check have changed and if the
unit should be sent in for service.
Yearly Service Evaluation
INFICON advises that the DataFID be sent to INFICON once a year for preventive maintenance
and calibration. Sending the unit to INFICON on a yearly basis allows the user to receive
upgrades to software, preventive maintenance, cleaning and calibration, along with a calibration
certificate. Contact INFICON Customer Service for information on ProCare
TM
Annual
Maintenance.
Maintenance Schedule Table
Item/Operation
Frequency/Usage
Battery Pack Charging
For 8 hours of use, charge 4 hours
Hydrogen Fuel Cylinder Filling
5 minutes for 10 hours of use (see Table 2
for cell charge times)
Inlet Filter Replacement
When dirty or once per week
Sintered Filter Replacement
Once every month in wet/concentrated
environments
Calibration - Daily
In condition unit will be used
Calibration - Monthly
Record readings. Notify INFICON to
discuss service options if inaccuracy
consistently over 20%
Waste Electrical and Electronic Equipment (WEEE)
EU regulations for the disposal of electric and electronic appliances that have been defined in the EU
Directive 2002/96/EC and in national laws became effective in August 2005 and apply to this device.
Common household appliances can be disposed of using special collecting and recycling facilities.
However, as the DataFID has not been registered for household usage, it must not be disposed of
through those means. The device can be returned to your national vendor/sales organization for
disposal. Please do not hesitate to contact your vendor/sales organization if there are any questions on
this issue.