beautypg.com

INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 48

background image

074-578-P1A

48

1/02/09 method 21 14:45

1/02/09 method 21 14::45

cancel view

view meth 21

When the user has a valid BACKGROUND reading on the display, simply press the ‘stop’ soft
key.

The maximum value measured between ‘start’ and ‘stop’ will be saved into the log

associated with the current Method 21 tag. The screen will change to that shown in Figure 24.

Figure 24. Method 21 Mode Component - Start

Now that the Background value has been successfully logged, the COMPONENT value must be
measured for the current tag. When the user is ready to begin measuring a COMPONENT
value, press the ‘start’ soft key. The screen will change to that shown in Figure 25.

Figure 25. Method 21 Mode Component - Stop

When the user has a valid COMPONEN

T reading on the display, press the ‘stop’ soft key. The

maximum value measured ‘start’ and ‘stop’ will be saved into the log associated with the current
Method 21 tag. The screen will change to that shown in Figure 26.

Figure 26. Method 21 Mode Component - Results

This view displays the Corrected value which is the maximum COMPONENT value minus the
BACKGROUND max value. It also displays the Leak Definition for the current tag and allows the
user to compare the Leak Definition with the Corrected value. Once the comparison is

54.3

ppm

Valve 12YJ5 H47

54.9

ppm

Valve 12YJ5

COMPONENT

cancel view

stop

start

Corrected:

52.6

ppm

Leak Def:

100

ppm

status

This manual is related to the following products: