Operational check list – INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual

Page 37

background image

074-578-P1A

37

If you will be in the field for more than one day you should include the following additional items:








Table 4. Additional Field Items

Operational Check List

Before beginning field work, set up and calibrate the DataFID for your particular application. Ensure the
instrument is in working order before heading into the field.

To prepare the DataFID for field work

1. Do not transport the DataFID with the hydrogen fuel cylinder attached to the DataFID. Remove
the hydrogen fuel cylinder from the DataFID and package it individually for transport.

2. Press the SETUP key and ensure the correct date and time are entered.

3. Program and calibrate all the Cal Memories that will be using. After calibration is complete,
sample the bag of calibration gas and the bag of zero air to ensure the DataFID has been
calibrated correctly.

4. If an averaging interval will be used, it is suggested to delete all events from the datalogger to
avoid confusion between different days’ data and to avoid running out of space in the datalogger.

5. If Method 21 monitoring will be performed, ensure all the Cal Memories are programmed and
calibrated. In addition, the monitoring schedule needs to be programmed as well.

Battery charger (Part No.A1201221)

Hydrogen Filling Station (Part No. A1201222)

Hydrogen fuel (Part No. MX754112)

Computer or PDA