INFICON DataFID Portable Flame Ionization Detector for Leak Detection and Repair (LDAR) User Manual
Page 74

074-578-P1A
74
9. The DataFID will automatically enter the span value. If there is a problem with the span gas
process, the DataFID will inform you, e.g. concentration for span gas too low.
10. Once the first span gas calibration is complete, the following screen will appear:
To go back to the first calibration step, press the back key, otherwise press Next.
11. This screen will appear. Keep this second value by pressing Next, or change it by pressing
New.
12. Attach span gas as instructed and press Next.
Calibrating Span Gas1
Perform Span 2
Calibrate next
Cancel
Back
Next
“next” to continue
OR ‘new’ to change
2,000 ppm
Cancel
New
Next
Attach
Span Gas 2
Cancel
Next