beautypg.com

Microfid ii troubleshooting – INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 80

background image

074-579-P1A

80

Cause: Ambient air is contaminated.

Action: If you are unsure about the quality of the ambient air, use a charcoal filter

or a supply of commercial zero grade air.

Cause: Hydrogen supply is contaminated

Action: Hydrogen may react with the carbon in the steel, inside the tank to

produce methane. This will only occur if the cylinder is in poor condition

and if the hydrogen has a high moisture content. Replace the hydrogen

tank. Empty and refill the MicroFID II hydrogen fuel cylinder with fresh hydrogen.

Fault: Signal from span gas is too small.

Cause: Span gas and zero air mixed up.

Action: Ensure calibration gas is used to calibrate the MicroFID II. Mark the calibration
and zero gas bags clearly.

Action: Ensure the span gas is of a reliable concentration.

Fault: Detector field voltage is low

Cause: Internal fault in the electronics.

Action: Contact the Service Department.

MicroFID II Troubleshooting

Problem: No instrument response detected, yet compounds are known to be present.

Cause: The MicroFID II has not been calibrated properly.

Action: Ensure calibration gas is of a reliable concentration and then calibrate

the instrument as outlined in Chapter 3. After the instrument has been

calibrated, sample the bag of calibration gas. A reading equivalent to the

calibration gas should be displayed. If not, contact the Service

Department.

Cause: Background contamination from the hydrogen.

Action: It is possible that the hydrogen has become contaminated and is contributing a
high background signal, and the hydrogen supply should be replaced with a new cylinder.
When ordering hydrogen, specify ultra high purity hydrogen, 99.999% pure. Empty Refill
the hydrogen fuel cylinder with the clean hydrogen supply.

Cause: There is an undetermined problem.

Action: Contact the Service Department.