location
14:45
1/02/09
meas
14:45
Location Mode
Figure 20. Location Mode Display
Location mode is identified by the 20 character Location string on the line below the current
TVOC value being displayed. Location mode will continuously display the instantaneous
concentration of total volatile compounds. Location mode also allows the user to manually
location and log readings. Location mode allows the user to datalog a reading by pressing the
accept key (soft keys 2 and 3 together).
Location strings are loaded into the MicroFID II via a PC using the INFICON ProComm
software. In Location mode, the soft keys are CANCEL, EDIT, and MENU.
CANCEL exits Location user mode and returns to the default Interval user mode.
EDIT selects “Modify Location numb”, “Edit current Location”, or “Insert new Location”.
MENU selects the MicroFID II
’s Main menu.
Interval Mode
Figure 21. Interval Mode Display
Interval displays the instantaneous readings. Interval mode also automatically stores these
readings in the MicroFID II
’s memory at a preset interval selected by the user. In Interval mode,
the soft keys are STATUS, VIEW, and MENU.
STATUS displays a read-only version of the important setup parameters.
VIEW displays a read-only view of the current data log.
MENU returns to the MicroFID II
’s Main menu while continuing to measure in the
background.
35.7
ppm
Location Route Description
CAL: Meth
cancel edit
menu
accept
45.6
ppm
CAL: Meth
status view
menu
sample