8. This screen will automatically appear: Press next once you attach the first span gas.

9. The MicroFID II will automatically enter the span value. If there is a problem with the span
gas process, the MicroFID II will inform you, e.g. concentration for span gas too low.

10. Once the first span gas calibration is complete, the following screen will appear:

To go back to the first calibration step, press the back key, otherwise press next.

11. This screen will appear. Keep this second value by pressing next, or change it by pressing
new.

Calibrating Span Gas1

Perform Span 2

Calibrate next

Cancel

Back

Next

Attach

span gas 1

Cancel

Next

“next” to continue

OR ‘new’ to change

2,000 ppm

Cancel

New

Next