INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 65

background image

074-579-P1A

65

edit span

connect span

spam level

6. The MicroFID II then asks for the span gas concentration. Enter the known concentration or
select New to enter a different span gas concentration and then connect the gas bag adapter to
the inlet.

Selecting New will prompt the user to reset the digits for the span gas value. The left and center
keys will change the corresponding values up or down while the right key will move the place
holder over to the right. If the place holder moves too far to the right, press the right arrow key
until the cursor moves back to the left to edit changes. Once the selection has been edited,
select the trigger key to accept the changes and continue to the Span segment of calibration.

7. Open the span gas (bag/regulator) and press Next

Enter new calibration

00500.0

done

Attach Span Gas

cancel

next

500 ppm

cancel

next