beautypg.com

INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 79

background image

074-579-P1A

79

Flame out may also occur when sampling enclosed or confined

spaces where vapors and gases cannot escape. Watch for

indications of increased flame height such as erratic readings or

sudden high concentrations followed by a flame out

fault.

If you will be using the MicroFID II in a highly contaminated area

where it is possible the oxygen content is below 10%, watch

for indications of reduced flame height such as lowered detection

limits or a flame out fault.

Cause: High concentrations of flammable gases (gases within their flammable range) are

present. High concentrations of flammable gases can act as an additional fuel

source. When this happens, the flame height may increase beyond the confines

of the combustion chamber. The hydrogen supply will then be cut off and the

flame will go out.

Action: Move to a location where there is an adequate supply of air and restart

the flame. See the information above. Watch for indications of increased

flame height such as erratic readings or sudden high concentrations

followed by a flame out fault.

Cause: Exhaust port is blocked.

Action: At low temperatures, water vapor, a by-product of the hydrogen

flame, may condense at the exhaust port. At sub-zero

temperatures, the water vapor will freeze and obstruct the

exhaust port. If the exhaust port becomes obstructed, pump

operation will be inhibited. Flame out may also result. Operate

the MicroFID II within the operating temperature range of 32°F (0°C) to 122°F (50°C).

Action: Replace the exhaust filter with a new or unblocked one. If this does not solve the

flame out issue, contact INFICON local service personnel.

Cause: Sample line is blocked.

Action: Ensure the sample line is not obstructed in any way

Cause: Inlet filter is plugged.

Action: Replace inlet filter. See Replacing Filters. See Replacing Inlet Filter section under
Routine Maintenance chapter.

Fault: Signal from zero gas is too high.

Cause: Contamination of sample line or fittings before the detector.

Action: Clean or replace the sample line or the inlet filter. See Replacing Inlet Filter
section under Routine Maintenance chapter

Cause: Span gas and zero air mixed up.

Action: Ensure clean gas is used to zero the MicroFID II. Mark the calibration and zero
gas bags clearly.