System alerts and alarms – INFICON MicroFID II Portable Flame Ionization Detector User Manual

Page 43

background image

074-579-P1A

43

MicroFID II Display Description

Flame Icon with line
through it

Flame has been extinguished and
needs to be relit

TVOC high level
alarm screen

TVOC concentration alarm
exceeded set value for selected
compound

Low Battery Icon

Low battery

Pump Error

Pump fault

Displayed value
blinking ON and
OFF

TVOC concentration over
instrument operating range

TVOC Display Max
Err

TVOC concentration over
instrument operating range

Hi Chamber temp

Chamber temperature too high

Recommend H2 Fill

Hydrogen cylinder has 10 hours of
use remaining

H2 Critically Low!

Hydrogen Cylinder has 5 hours of
use remaining

Sending log data to
the PC

Instrument sending log data to PC

Cal Zero Error

Zero value too high during
calibration

Span Gas Error

Span gas too low during
calibration

Table 5. System Alerts

System Alerts and Alarms

While operating the instrument, system alerts can occur. To accurately identify the source of the
alarm, each type of alarm has been given a unique status.

In addition to the status, the MicroFID II also has an audio alarm and an alarm LED. A soft key
is

used for acknowledging alarms and is named “Accept”. To clear the alarm, press the “Accept”

key. Once acknowledged, the alarm indicators are cleared. The alarm status will remain until the
alarm condition clears.

The MicroFID II updates the instantaneous concentration once every second. Following every
update, the instantaneous concentration is compared to the selected compounds high alarm
level, and if exceeded, an alarm is triggered.

During calibration, all alarms are disabled. Once the calibration is complete, the alarms are re-
enabled.