Faults and troubleshooting

1.01

Rotacool mini

33

6 Faults and troubleshooting

Fault/Message

Cause

Remedy

Device cannot be switched on.

Power cable not connected to the
mains supply

 Plug power cable into the

mains supply.

Fuses defective

 Replace fuse (see chapter

7.2, page 34)

No circulation of cooling liquid

Main power switch defective

 Contact service

Level of cooling liquid too low

 Amount of cooling liquid too

low, add cooling liquid (see
chapter 5.4.3, page 30)

Fuses defective

 Replace fuse (see chapter

7.2, page 34)

Faulty temperature sensor

 Contact service

Does not cool

Air intake of chiller contaminated
or obstructed

 Clean rear panel (see chap-

ter 5.4, page 28)

Table 6-1: Troubleshooting


7 Maintenance, cleaning, service

7.1 Maintenance

The liquefier is located on the base panel of the Rotacool mini casing. Cooling
air is drawn through it from below. Avoid light particles in the air intake on the
bottom of the device. The base panel should be cleaned from time to time.

 Remove the circulating chiller power cable and empty the device entirely

(see chapter 8.1.2 page 36), remove the cooling hoses on the rear of the
device and tilt the casing over onto its side.

 You can clean the air intake on the base panel with a vacuum cleaner or

brush. However, refrain from using any sharp-edged or pointed objects.
The circulating chiller should be placed upright again immediately after
cleaning. Before reactivating, wait about 20 minutes, so the oil in the com-
pressor can collect again.