Initial operation, 4 initial operation – Heidolph Rotacool Mini User Manual
Page 30

Setup and initial use
28
Rotacool mini
1.01
Distance from walls and ceiling to ensure adequate ventilation (dissipation
of waste heat, fresh air supply for thermostat and working room), min. 20
cm
Ambient air temperature of a minimum of 5°C up to maximum of 32°C; it is
mandatory to maintain the correct ambient conditions to ensure fault-free
operation.
Highest relative humidity is 80% for temperatures up to 32°C;
Short distance to supply connections
The device must not be set up in such a way that access to the mains
connection switch is hindered or obstructed;
Fluctuation in supply voltage must not exceed ±10% of the nominal volt-
age;
Transient surges, as commonly occur in the mains network;
Applicable pollution level: 2;
Overvoltage class II.
Avoid setting up the circulating chiller in a corrosive atmosphere.
Choose a level, firm base.
5.4 Initial operation
Fig. 5-1:
Rotacool mini Front view
1 Device switch
2 Fill level indicator; overflow hose; drain hose
3 Filler connector